Monarch Instrument ACT-1B Series User Manual

Page 6

Advertising
background image

3

USB Programming Cable and Software

The 3.5 mm connection “hole” in the center bottom of the rear panel is for the
optional USB Programming Cable (p/n 6180-031), which comes with Windows ™
compatible PM (Panel Meter) Remote Software on CD. The cable and software
combination allows the user to configure: operation mode, analog output scaling,
decimal places, display update rate and pulses per input. The user can also view real
time data in digital format and/or through Microsoft

®

Excel. Software instructions are

included on the software CD.

SENSOR CONNECTIONS

A speed sensor (not included) is connected to the terminals under the section
labeled INPUT on the rear panel. Refer to Figures 2 and 3.

Connections and their functions are as follows:

+VA

Positive +12 Vdc to provide power to optical, laser, infrared
or amplified magnetic sensors. Maximum load is 75 mA dc.
Optionally 5 or 24 Vdc supply may be ordered.

PX+

Positive +12 Vdc supply for use with two-wire proximity
sensors. Maximum load for proper operation with two-wire
sensors is 25 mA.

SIG

Positive input signal from the speed sensor. Accepts TTL

pulses or ac signals, unipolar and bipolar, from 1.5 Vac to 50
Vac. (Contact the factory for increased sensitivity.) Connect
the signal wire from three-wire sensors or the positive side of
two-wire magnetic sensors to this terminal. Typical input
impedance is 10 Kohms.

COM Common or Negative connection for both signal and power

from most sensors.

Refer to Figure 3 for connection of Monarch standard sensors. The
connections are typical for these types of sensors.

This product is designed to be safe for indoor use under the following conditions
(per IEC61010-1):

Installation Category II

per IEC 664

Pollution Degree Level II

per IEC 664

Operating Temperature

32-122 °F [0-50 °C]

Humidity

Maximum relative humidity 80% for temperature up to 88 °F
[31 °C] decreasing linearly to 50% relative humidity at
104 °F [40 °C]

OPTIONS

IO:

4 to 20 mA current output

AO:

0 to 5 Vdc analog output

NOTE: Full scale RPM must be specified for the above

options when ordering unless the USB
Programming Cable and PM Remote Software is
ordered as a separate item.

PO:

0 to 5 V TTL compatible pulse output

NOTE: Pulses out per revolution equal pulses in per

revolution.

CAL-N.I.S.T.

N.I.S.T. Traceable Certificate of Calibration

8

Advertising