Output options, Specifications, Figure 3 sensor connections – Monarch Instrument ACT-1B Series User Manual

Page 7

Advertising
background image

4

Figure 3 Sensor Connections

OUTPUT OPTIONS

The ACT-1B may be equipped with either a Current Output (IO) or an
Analog Output (AO), and/or a TTL Pulse Output (PO).

NOTE:

Full scale RPM settings must have been specified when ordered
or may be user programmed using the optional USB Programming
Cable and PM Remote Software.

The Current or Analog Outputs are connected to the terminals in the section
labeled ANALOG OUTPUT OPTION on the rear panel. The actual output
is marked. The Pulse Output is connected to the terminals under the
section labeled OUT on the rear panel.

CAUTION:

The IO or AO COM may NOT be isolated from the other
COM connections.

Shield to
COM

Shield to
COM

PX+

+VA

COM

SIG

INPUT

PX+

+VA

SIG

COM

INPUT

PX+

+VA

SIG

COM

INPUT

BLACK

WHITE

BROWN

BLUE

SIG

INPUT

+VA

PX+

COM

BROWN

BLACK

BLUE

AC or TTL

SOURCE

M-190W

MAGNETIC

SENSOR

P5-11

PROXIMITY

SENSOR

ROS-W - OPTICAL SENSOR (SHOWN)

R0LS-W - REMOTE LASER SENSOR

IRS-W - INFRARED SENSOR

MT-190W - AMPLIFIED MAGNETIC SENSOR

Link SIG to
COM

7

SPECIFICATIONS

Range:

5 to 99,999 RPM

Accuracy:

±1 RPM or 0.005% of reading

Resolution:

1 RPM (user programmable to 0.0001*)

Display:

5 digit, 0.56” (14 mm) high red LED

Display Update: Twice per second above 120 RPM (user programmable

to 0.5, 1, or 1.5 second*)

Dimensions:

1/8 DIN by 4.5” (114 mm) deep

Power Supply:

Standard: 100 - 240 Vac ±10%, 50/60 Hz, 5VA
Optional: 12 or 24 Vdc ±10%, Isolated, 5 Watts

Inputs:

Universal input for optical, proximity, two-wire or three-
wire magnetic, infrared or laser sensors
TTL input or 2 Vac to 50 Vac peak to peak input
Standard input is 1 pulse per revolution. (user
programmable from 1 to 999*)

Sensor Excitation: 12 Vdc at 20 mA for proximity sensors, 12 Vdc at 75 mA

for all other sensors, Optional 5 or 24 Vdc available

Recommended Sensors: Optical

- Monarch ROS-W

Proximity - Monarch P5-11
Magnetic - Monarch M-190W or MT-190W
Infrared

- Monarch IRS-W

Laser

- Monarch ROLS-W

IO Option:

4 to 20 mA out, 16 bit resolution 12Vdc compliance voltage. See
page 7 for maximum load calculation. Full scale RPM settings as
specified when ordered. (user programmable*)

AO Option: 0 to 5 Vdc out, 5 mA 16 bit resolution. Full scale RPM settings

as specified when ordered. (user programmable*)

PO Option: 0 to 5 V TTL pulse, non-inverting, one pulse out for each

pulse in. (Positive or negative out programmable*)

* Requires optional USB Programming Cable and PM Remote Software