11

To set a value greater than 99999, press the RECALL button causing all the decimal points to light,
indicating that the displayed value must be multiplied by 10. Thus, any value up to 999990 can be set.

NOTE: If the most significant digit (that on the extreme left) is zero, you cannot set the times ten

mode.

To escape from the EDIT Mode at any time without updating the value, simply press the RESET
button.

To exit from the EDIT Mode and update the value, press and hold down the SET button and then
press the RESET button. Always ensure that the value entered is within the constraints laid down or
you will get an Err 7 message, and the value will not be updated. Err 5 will be displayed if you try to
set a limit that is not enabled. If the value is acceptable the unit will display donE.

NOTE: Any time the limit value is changed, the instrument will automatically recalculate the hysteresis

at 5%. Refer to Sections 6.8 and 6.9.

5.1

Edit Mode Button Summary (Changing Digits)

When in the EDIT Mode as described above, the keypad buttons have the following functions:

SET

- changes the digit being edited

RESET

- aborts the edit process

UP (

∧∧∧∧∧)

- increases the digit by 1

DOWN (

∨∨∨∨∨)

- decreases the digit by 1

RECALL

- shifts the decimal point

SET + RESET

- saves data and exits

SET + DOWN (

∨∨∨∨∨) - toggles a minus sign in the left most position on the display.

6.0

USING THE MENU

To enter the MENU Mode, the user must use a combination of three keys. Press and hold down the
SET, RESET and RECALL buttons. Press the SET button first if you do not wish to reset the limits,
if they are present. The instrument will enter the MENU Mode, and the display will show tyPE.

There are a number of menu selections, some of which may not be seen depending on the option
ordered, the options set, and the general status of the instrument. See page 25 for an overview of the
menu options. Below is a list of the menu items with the function(s) they perform. Use the UP (

∧∧∧∧∧) and

DOWN (

∨∨∨∨∨) buttons to step through the main menu options.

To exit from the MENU Mode press the RESET button.

6.1

TYPE (Mode)

Sets the mode of operation. This menu item has a number of sub items. Once selected with the
SET button, the instrument will display its current mode. To exit back to the main menu without
changing the mode, press the RESET button. If you wish to change the operating mode, you can
view the options using the UP (

∧∧∧∧∧) button. The available options are RPM, FREQuency, SCALE,

SECAP (Single Event Capture), ROC (Rate of Change) and TOTALizing. These functions are
described in Section 1. When the desired mode is displayed, press the SET button. The unit will
display DONE and return to the main menu.

6.2

Point (Decimal)

Sets the decimal point to either AUTO or SET (FIXED). To enter this menu item press the SET
button. The unit will display the current decimal point setting. To exit back to the main menu
without changing the mode, press the RESET button. To select the other mode, press the UP (

∧∧∧∧∧)

button which will toggle between the two modes. Press the SET button to change the decimal
point setting to the option displayed and exit back to the main menu.

12

6.3

SET 1 (Alarm 1)

NOTE: This menu option will not be seen on an ACT-2A.

Used to set the LIMIT 1 type. This menu item has a number of sub items. HI or LO determine
whether the limit is high or low, LATCH or RESET determine whether the limit will be latching
or non latching, and LOC or NOLOC determine whether the lock out feature will operate for the
LO limit setting.

Select this item with the SET button. You can exit back to the main menu at any time without
changing the current settings by using the RESET button. Once selected, the display will show
the current setting of the limit type. If no limit is set it will show OFF. You may cycle through
the three choices (HI, LO, OFF) by using the UP (

∧∧∧∧∧) button. Press the SET button when the

desired limit type is displayed.

The unit will then display LATCH or RESET depending on how it is currently set. Again, use the
UP (

∧∧∧∧∧) button to change the type. Pressing RESET at this time will not alter the limit. Use the

SET button to continue. If you set the limit to HI, you will return to the main menu. If you
selected LO, you will have to choose the LOC or NOLOC option as described above.

NOTE: Setting the limit type will force a recalculation of the hysteresis to 5%.

6.4

SET 2 (Alarm 2)

Same as SET 1 but for LIMIT 2.

6.5

SERNO (Serial Number

Used to view the serial number of the unit. Use the SET button to view the serial number and the
RESET button to return to the main menu.

6.6

SCALE (Scale Factor)

Used to set the scale factor for the Scale Mode of operation. This menu option will not be seen
if the unit is not set in the Scale Mode. Enter this menu item with the SET button. The current
scale factor will be displayed. To exit to the main menu without changing the value press the
RESET button or else change the value in the same way that the limits are changed, as in Section
5.0. To update the scale factor, press the SET and RESET buttons. The permissible values for
the scale factor are from 0.0001 to 9999.9.

6.7

TOTAL (Totalizer Scale Factor)

Used to set the scale factor for the Totalizer mode of operation. This scale factor is independent
of the SCALE mode scale factor. It will not be seen if the unit is not set in the TOTAL mode, using
the TYPE menu option. Enter this menu item with the SET button. The current Scale Factor will
be displayed. To exit to the main menu without changing the value press the RESET button, else
change the value in the same way that the limits are changed, as in Section 5.0 and 5.1. To update
the scale factor, press the SET and RESET buttons. The permissible values for the scale factor
are from 0.0001 to 9999.9.

6.8

HYST1 (Hysteresis Limit 1)

NOTE: This menu option will not be seen on an ACT-2A or if the limit has been set as non-

latching.

Used to set the limit hysteresis as a percentage. Once selected with the SET button, the existing
hysteresis value will be displayed. It may be changed in the same manner as setting a limit. Refer
to Section 5.0 and 5.1. The permissible values for hysteresis are 0.0001 to 99.999%. The RESET
button escapes back to the main menu, while the SET + RESET buttons pushed together will
update the hysteresis value.