4 medical vacuum systems glossary – Amico NFPA Triplex RVL Modular Stack Mount User Manual

Page 10

Advertising
background image

Notes on the Source Sizing Guide

These tables represent standard systems and configurations and do not represent all configurations.

In particular:

Envelope dimensions do not include maintenance space. Amico Source recommends 36 in. (92 cm) on all sides but
less space may be possible with an understanding of the system maintenance requirements. In all cases the National
Electrical Code requires a minimum clearance of 36 in. (92 cm) in front of the control cabinet.

Contact your Local Amico Source representative should you require fitting the systems in smaller spaces.

All systems listed have the standard receivers. Larger receivers usually change the envelope dimensions.
The modular system’s envelope dimensions shown represent a typical arrangement for minimum floor space. How-
ever, modular systems are easily located wherever appropriate, which will change the dimensional requirements.
Envelope dimensions for modular systems should be used only for informational purposes.
More details on all dimensions will be found on the spec sheets for the system selected and the spec sheet should
be consulted when doing final layout.

8

Definitions

ACFM

Actual Cubic Feet per Minute is an expression of actual air volume, generally

corrected for and reference to a particular pressure.

Displacement

Theoretical physical volume of the air in the pump chamber based upon 100%

pumping efficiency, with no allowances made for heat, friction, clearances, or

other losses in the compression cycle.

Duplex System

Systems comprised of two (2) pumps, each rated for 100% Peak Calculated Load

(PCL)

pipeline

system.

Intermittent Duty

Reference to pumps not capable of operating continuously (i.e. Pump design

requires periodic shut-down for cooling or oil transfer).

NTP

“Normal Temperature and Pressure”, generally accepted as 70˚F (21˚C),

14.969 psi/29.92” Hg barometric pressure, and 36% relative humidity.

Simultaneous Demand

Operating reference to a condition where all pumps (Lead and Lag pumps) run

simultaneously to satisfy demand in excess of lead pump(s) capability.

Timed Alternation

Lead and lag pumps alternate on a timed basis. In medical vacuum applications,

all systems should be able to operate on a timed alternation bases to ensure

equal wear of the vacuum pumps.

Source End Vacuum

The vacuum (negative pressure) level required at the vacuum system source in

order to provide the vacuum level at terminal units (outlets) required by

the specific suction regulators utilized in the facility.

Peak Calculated Load (PCL)

The maximum estimated demand a medical facility will require of a Medical

Vacuum

System.

Calculated at SCFM at 483mmHg (20” Hg).

LPM

Abbreviation for Liters per minute; a measure of the flow rate of a gas.

SCFM

Standard Cubic Feet Per Minute is an expression of air at NTP.

Triplex Systems

Medical vacuum system with three pumps (3), each sized for 50% PCL.

Quadruplex System

Medical vacuum system with four pumps (4), each sized for 33% PCL.

Continuous Duty

Operational reference to pumps operating 24 hours a day, continuously.

Continuous Duty Rated

Vacuum systems which can operate continuously (24 hours per day) if

necessary but normally only operate on demand.

7.4 Medical Vacuum Systems Glossary

Amico Source Equipment

Advertising