Psi / kpa / bar selection – Amico Valve Alarm Combo Unit User Manual

Page 14

Advertising
background image

Factory Default - PSI

For PSI mode, set the switch #4 to the ON position. The LED PSI indicator located next to the

gas pressure reading will illuminate.

For kPa mode, set the switch #4 to the OFF position and switch #9 to the ON position. The LED

kPa indicator located next to the gas pressure reading will illuminate.

For BAR set the switch #4 to the OFF and the switch #9 to the OFF position. The LED kPa

indicator located next to the gas pressure reading will illuminate. (There is no separate indicator

for BAR).

Vacuum Only

Vacuum alarm set-point adjustment
Factory Default:
High = 30"Hg, Low = 12"Hg
Repeat time = 30 min.

During programming, the “Trend Bar” will flash!

1. Set switch #6, #7 and #8 to the ON position.

2. The LED will display (HI-), followed by the current set point. Indicating the system is ready to

accept a new High set point. Do not adjust this set point since the High set point is not used.

3. Set switch #7 to the OFF position.

4. The LED will display (LO-), followed by the current set point. Indicating the system is ready

to accept a new Low set point. Adjust set point, using the “UP” and “DOWN” push buttons,

to the desired value.

5. Set switch #8 to the OFF position.

6. The LED will display (I-I-), followed by the current set point. Indicating the system is ready

to accept a new Repeat Time set point. Adjust set point using the “UP” and “DOWN” push

buttons, to the desired value. [(Display dd=Disabled) Range from 1 to 60 Minutes].

7. Set switch #6 to the OFF position.

When you have completed step #7, the display module will automatically go into “RESET” mode.

This will store the data that you entered.

PSI

KPA

BAR

PSI

KPA

BAR

PSI / kPa / BAR Selection

14

Amico Pipeline

Advertising