Amico Valve Alarm Combo Unit User Manual

Page 19

Advertising
background image

Symptom

Cause

Corrective Action

Gas reading incorrect

Loose connection of DISS fittings

Ensure that the sensor module is properly

connected to the DISS demand check-valve

Sensor module is not properly wired to the

display module

Ensure that the sensor module is properly

wired to the display module by using wiring

diagram in Appendix C or D

Defective sensor or requires calibration

Replace the sensor module

The ribbon cable is not properly connected

to the display module

Pull out the ribbon cable and connect

it back in again, while ensuring that it is

seated properly

Defective display module

Replace the display module

System

Cause

Corrective Action

Page

E01

No sensor is connected.

Connect a sensor

28/29

E02

Sensor and Display Module mismatched.

Ensure that the Sensor and Display Module are for the

same gas

16

E03

The High set-point was set below the Low set-point

or vice versa.

Recalibrate the High and Low setpoint to proper

values

13/14

E04

Incorrect type of Sensor connected

(e.g. 250 psi sensor on a 100 psi range)

Connect the correct Sensor to the matching Display

Module

E06

Cable between the sensor and display module

shorted out or reversed polarity

Reverse polarity or replace cable if defective

28/29

E07

Out of calibration / Sensor not reading gas

Replace the sensor module

28/29

Error Code Messages on the Display Module

Troubleshooting Guide

www.amico.com

19