Troubleshooting, Factory default setting gas – Amico Alert-3 LCD Alarm User Manual

Page 18

Advertising
background image

18

Amico Pipeline

Troubleshooting

FACTORY DEFAULT SETTING GAS

Mid Pressure

Hi

= 60 psi

Low

= 40 psi

Vacuum

Hi

= 32 inHg

Low

= 12 inHg

High Pressure

Hi

= 195 psi

Low

= 140 psi

Symptom

Cause

Corrective Action

Gas reading incorrect

Loose connection of DISS fittings

Ensure that the sensor module is properly

connected to the DISS demand check-

valve

Sensor module is not properly wired to

the display module

Ensure that the sensor module is properly

wired to the LCD module by using wiring

diagram in Appendix D or Appendix E

Defective sensor or requires calibration

Replace the sensor module

Defective LCD display module

Replace the LCD display module

Display shows “NO SENSORS”

No sensor(s) are connected to the LCD

board

Make sure sensor module(s) are

connected to LCD board (see Appendix D

and Appendix E).

Program not set up

Press setup and select button to program

all connected sensors (see LCD Display

Set-Up in Section 9)

Faulty sensors

Replace sensors