User responsibility, Installation guide, Warning – Amico Alert-3 LCD Alarm User Manual

Page 8: Caution, Danger, Danger and

Advertising
background image

8

Amico Pipeline

A m i c o M i c r o p r o c e s s o r B a s e d A l a r m

P a g e : 1 2

CAUTION: The microprocessor circuitry on the ALERT-2 alarm con-

tains sophisticated integrated semiconductors. If it becomes necessary

to remove a module, PLEASE hold the boards by the edges.

DO NOT

TOUCH any of the components on the board. Static discharge can

cause the modules to malfunction, or become damaged.

SENSOR

LOCAL (In the Back Box)

1.

Locate the gas specific

sensor module to be in-

stalled.

2.

In the back box, there are

colour coded gas labels

located under the DISS

Demand check valves. Each

label identifies where each

sensor module is to be

placed.

3.

The sensor module contains

a gas specific DISS fitting.

Push the sensor module

hex-nut and nipple adapter

up into the demand check-

valve. With a wrench, tighten the nut so that it makes a good seal.

NOTE: Pressure on sensors are not to exceed

250psi

for Pressure sensors and

30”

for Vacuum sensors

REMOTE (Outside the Back Box)

1.

Connect a Tee (supplied

by others) to the pipeline

with a 1/4" NPT female

connection that will accept

the DISS Demand check-

valve.

2.

Locate the gas specific

sensor module to be

installed.

3.

Thread the DISS Demand

check-valve into the

correct gas pipe line.

4.

The sensor module contains a gas specific DISS fitting. Push the

sensor module hex-nut and nipple adapter up into the demand

check-valve. With a wrench, tighten the nut so that it makes a

good seal.

B: REMOTE (outside the back box)

i. Connect a tee (supplied by others) to the pipeline with a 1/4” NPT female

connection that will accept the DISS Demand Check Valve.

ii. Locate the gas-specific sensor module to be installed.

iii. Thread the DISS Demand Check Valve into the correct gas pipeline.

iv. The sensor module contains a gas-specific DISS fitting. Push the sensor

module hex-nut and nipple adapter into the demand check-valve. With a

wrench, tighten the nut so that it makes a good seal.

STEP 5: FRAME/MODULE ASSEMBLY

i. Remove the frame/module assembly from its protective box.

ii. Remove the corner screws from the front frame section (4 screws).

iii. Attach the LCD module to the back box assembly by using flat head screws

(provided with frame in a plastic bag) to the hinge located on the back box.

iv. Attach the frame wire with 2 dome head screws (provided with frame in a

plastic bag). This will allow the frame assembly and back box to be fastened

securely together.

CAUTION: The microprocessor circuitry on the ALERT-3 alarm contains sophisticated integrated

semiconductors. If it becomes necessary to remove LCD circuit board, PLEASE hold the boards by the

edges. DO NOT TOUCH any of the components on the board. Static discharge can cause the modules

to malfunction or become damaged.

A m i c o M i c r o p r o c e s s o r B a s e d A l a r m

P a g e : 4

USER RESPONSIBILITY

The information contained in this Installation and Operation Maintenance

Manual, pertains only to the ALERT-2 microprocessor based digital alarm.

This product will perform to conformity with the descriptions contained in this

manual, when assembled, operated, maintained and serviced in accordance

with the installation instructions provided.
The alarm

must be checked periodically. Parts that are broken, missing,

worn, distorted or contaminated,

must be replaced immediately. Should such

repair or replacement become necessary, please contact Amico Corporation

or their distributors.
All alarms should not be repaired, or altered without prior written or verbal

approval of Amico Corporation or it’s distributors. Failure to comply will void

all warranty on the alarm.
Statements in this manual preceded by the words

WARNING

,

CAUTION

,

DANGER and

NOTE

are of special significance. Please read these sections

carefully.

WARNING:

denotes steps which can prevent injury.

CAUTION:

denotes steps which can prevent damage to equipment.

DANGER:

denotes steps which can prevent electrical shock to equipment

or to prevent serious injury and/or death.

Installation Guide