Installation guide – Amico Alert-3 LCD Alarm User Manual

Page 7

Advertising
background image

www.amico.com

7

Installation Guide

STEP 1: THE ALARM BOX

Install the back-box to the studs of the wall at the desired height. Ensure that the box is securely in place. The mounting

brackets are adjustable to suit the thickness of the wall. MAKE SURE the box is parallel, squared and flush with the

finished wall surface to ensure that the frame assembly will fit properly.

STEP 2: FOR LOCAL SENSOR ONLY

If the sensors are to be mounted locally (inside the back box), the pipe stubs must be connected to the pipeline. Using

silver-brazing techniques, connect each pipe stub to its appropriate gas or vacuum while ensuring that the bottom of

the pipe stub is wrapped with a damp cloth. BE CAREFUL not to damage the DISS check-valve by overheating the lower

portion of the copper pipe. When the brazing of pipe stubs has been completed, the system can be pressure tested.

STEP 3: STANDING PRESSURE TEST

Perform a standing pressure test on the piping system as per NFPA-99 “Health Care Facilities.” Inspect all joints for leaks

and make certain each gas is piped to a correspondingly labeled gas service.

STEP 4: SENSOR

LOCAL (inside the back box)

i. Locate the gas-specific sensor module to be installed.

ii. In the back box, there are color coded gas labels located under the DISS Demand Check Valves. Each label

identifies where each sensor module is to be placed.

iii. The sensor module contains a gas-specific DISS fitting. Push the sensor module hex-nut

and nipple adapter up into the demand check-valve. With a wrench, tighten the nut so

that it makes a good seal.

Note: Pressure on sensors are not to exceed 250 psi for pressure sensors and 30 inHg for vacuum

sensors.

Alert-3 sensor operating pressure range:

Mid Pressure (0 to 99 psi)

-

Oxygen, Medical Air, Nitrous Oxide, Carbon Dioxide

High Pressure (0 to 249 psi) -

Nitrogen, Instrument Air

Vacuum

(0 to 30 inHg) -

Vacuum, WAGD, AGSS