Testing for leakage – Amico Dome Loaded Manifold CSA User Manual

Page 18

Advertising
background image

18

Amico Pipeline

Testing for Leakage

Precautions

The Amico Automatic Gas Manifolds are designed in accordance with the CSA Standard.

There are three categories of Amico manifolds depending upon the delivery pressure. The following gas types are

available for each delivery pressure:

55 PSI Delivery Pressure

• Oxygen, Nitrous Oxide, Medical Air, Carbon Dioxide, Helium, Argon

100 PSI Delivery Pressure

• Oxygen, Medical Air, Carbon Dioxide

170 PSI Delivery Pressure

• Nitrogen, Instrument Air

Delivery (Line) Pressure - (see NOTE above)

Delivery (Line) Pressure

55 PSI

100 PSI

170 PSI

Intermediate Pressure - Ready Bank

100 PSI

200 PSI

200 PSI

Dome Bias Pressure

55 PSI **

50 PSI

50 PSI

Intermediate Pressure - In Use Bank

150 ± 10 PSI*

250 ± 10 PSI

250 ± 10 PSI

Intermediate Relief Valve

200 PSI

350 PSI

350 PSI

Line Regulator Relief Valve

75 PSI

150 PSI

225 PSI

Maximum Inlet Pressure

3000 PSI

3000 PSI

3000 PSI

* The intermediate pressure valve of the “In Use” bank is dependent upon the dome bias pressure. Variations from the 55

PSI delivery pressure will affect the intermediate pressure reading.

** Same as delivery pressure.