Control board replacement – Amico Dome Loaded Manifold CSA User Manual
Page 28
28
Amico Pipeline
Control Board Replacement
If necessary, the control board replacement can be performed while the manifold is in service. However, this should only
be done by qualified technicians experienced in servicing medical equipment.
1. Remove two top screws at the base of the display board bracket and tilt the display board assembly backwards.
2. Disconnect power supply.
3. Remove the screws located on the circuit board, which detaches it from the front plate. Note orientation of the
old circuit board and remove the wiring one at a time and connect each wire to the appropriate terminal on the
new board.
4. Install the circuit board and fasten the screws to secure it. Tilt the circuit board to its original position and then
fasten the two screws at the base.
5. Reconnect power supply to the manifold cabinet.
6. Check the correct operation of the circuit board by simulating a bank switch over a couple of times.