Amico Pipeline

Control Board Replacement

If necessary, the control board replacement can be performed while the manifold is in service. However, this should only
be done by qualified technicians experienced in servicing medical equipment.

1. Remove two top screws at the base of the display board bracket and tilt the display board assembly backwards.

2. Disconnect power supply.

3. Remove the screws located on the circuit board, which detaches it from the front plate. Note orientation of the

old circuit board and remove the wiring one at a time and connect each wire to the appropriate terminal on the

new board.

4. Install the circuit board and fasten the screws to secure it. Tilt the circuit board to its original position and then

fasten the two screws at the base.

5. Reconnect power supply to the manifold cabinet.

6. Check the correct operation of the circuit board by simulating a bank switch over a couple of times.