Bio-Rad Image Lab™ Software User Manual

Page 178

Advertising
background image

A

| Maintenance and Specifications

178

| Gel Doc EZ Imaging System with Image Lab Software

5.

Carefully slide out the lamp assembly. Although only one lamp bulb might be
out, it is recommended that all lamps be replaced to ensure even illumination
and to reduce the need to access the lamps again.

6.

Replace all lamps with UV-B fluorescent lamps (Bio-Rad part catalog #900-
0217). Make sure each lamp is properly seated in its holder.

7.

After replacing all the lamps, carefully slide the lamp assembly back into the
instrument. Press on the side of the lamp assembly to the right of the screw
hold-down tab, to make sure it is fully seated into the instrument. If the lamp
assembly is not fully seated, the following error message appears the next time
it communicates with Image Lab software: Light tray not detected.