Setting Up a Default Protocol

User Guide

| 75

If you choose Faint Bands, exposure time is longer, making faint bands

more visible. More prominent bands might be overexposed.

Select manual exposure to override automated imaging. Exposure time

can range from 0.001 to 10 sec.

After a gel is imaged with automatic exposure optimization, the previous

exposure time displays in the protocol as a guide so you can manually adjust

the exposure time if necessary.

You can also view the exposure time of the image later, in the Image Info

window (see

Image Info on page

96

for more information).

3.

Set the Display Options

Highlight saturated pixels — select the Highlight saturated pixels checkbox to

see any saturated pixels in red, which shows how much of the gel image is

saturated. Saturated pixels are beyond the maximum quantifiable range of the

imaging system. This option can be changed later by selecting View > Image

Transform.

Image Color — select color choice to display the sample image. Viewing the

image with a different color scheme can make it easier to see all of the

elements. See

Image Colors on page

94

for the color choices.