UV or UV/Vis Detector

Instrument Guide

| 257

Sample trace values
are lower than
expected.

The lamps might be nearing the
end of their usable life.

Check the lamp usage values in
the System Information dialog
box. From the System Control
tab in ChromLab open File >
System Information > Detectors
tab.

If the lamp usage values are
close to their limits
(approximately 2,000 hours for
the deuterium and tungsten
lamps; approximately 5,000
hours for LEDs), change the
lamps.

For single-wavelength LEDs,
see

Replacing the Single-

Wavelength UV Detector LED
on page

201

.

For multi-wavelength lamps,
see

Replacing the Multi-

Wavelength UV/Vis Detector
Lamps on page

209

.

Symptom

Possible Cause

Possible Solution