Bio-Rad ChromLab™ Software User Manual

Page 69

Advertising
background image

Detectors

Instrument Guide

| 67

Monitors UV absorbance one wavelength at a time

255 nm

280 nm (default)

Tip:

Use ChromLab to change the absorbance wavelength values. For

more information, see the

NGC Chromatography Systems and ChromLab

Software User Guide.

UV absorbance range: 0–3,000 mAU

UV linearity: ±5%

Three interchangeable flow cells with path lengths of 2 mm, 5 mm (default),

and 10 mm; see Table

4 on page

69 for specific information

Note:

Use the preparative 2 mm flow cell to measure high concentrations

of protein. The path lengths of the analytical 5

mm and 10 mm flow cells

are longer and are appropriate for measuring lower concentrations of

protein.

For information about changing the flow cells, see

Replacing the UV Flow Cell

on page

196

.