EXPLANATION OF FUNCTIONS

The DFI 200 setup and calibration functions are configured through a push button sequence.

Two levels of access are provided for setting up and calibrating namely:

mode (simple push button sequence) allows access to calibration channel selection, display

rounding, decimal point and digital filter settings only.

mode (power up sequence plus push button sequence) allows access to all functions including

calibration parameters.

Once

or

mode has been entered you can step through the functions, by pressing and releasing the F

push button, until the required function is reached. Changes to functions are made by pressing the ^ or v push
button (in some cases both simultaneously) when the required function is reached. Any changes to a function will
not be saved until the F button is pressed to accept the change.

4.1 Calibration internal memory selection

Display:

Range: CH1,

,

or

Default Value: CH1

Selects one of the three possible calibration internal memories (CH1,

,

, also

if the optional

external memory chip is used - see chapter

4.24

). The

function allows the instrument to be calibrated to up

to 3 different load cells and hold the calibration values in internal memory. Alternatively 3 different calibration