Cooper Instruments & Systems DFI 200 Hand Held Indicator User Manual

Page 6

Advertising
background image

CF 112

6

Version 3.1-0

Default Value:

Displays and sets the decimal point place. By pressing the

or

pushbuttons the decimal point position may

be set. The display will indicate as follows:

(no decimal point),

(1 decimal point place),

(2

decimal point places) or

(3 decimal point places).

4.5 Digital filter

Display:

Range:

to

Default Value:


Displays and sets the digital filter value. Digital filtering uses a weighted average method of determining the display
value and is used for reducing display value variation due to short term interference. The digital filter range is

selectable from

to

, where

= none and

= most filtering. Use

or

at the

function to alter

the filter level if required. Note that the higher the filter setting the longer the display may take to reach its final value
when the input is changed, similarly the relay operation and any output options will be slowed down when the filter

setting is increased. To set the digital filter value go to the

function and use the

or

push buttons to

set the required value then press

to accept this selection.


4.6 mV/V input range

Display:

Range:

or

Default Value:


Displays and sets the mV/V range to suit the transducer useable range. Choose the value equal to or the next
higher value to the mV/V output of the transducer. This selection sets the input range for the A/D converter. If too
low a range is selected a

error message will be seen on the display. If too high a range is selected

the full resolution capability will not be used and problems with calibration can result.

4.7 Sample rate

Display:

Range:

Default Value:


Sample rate for ADC - Displays and sets the ADC sample rate from 5 to 100 samples per second. The faster
sample rates can be utilized in features such as peak hold and peak/valley memory.

4.8 User pushbutton function

Display:

Range:

Default Value:

In normal measurement mode the User push button can be programmed to operate in one of the modes described
below. In CAL or FUNC modes pressing the User button will cause the instrument to immediately return to normal