Cooper Instruments & Systems TT03C Digital Cap Torque Gage User Manual

Page 15

Advertising
background image

Series TT03 Digital Torque Gauges

User’s Guide


14

If failed:

5. The following screen appears after the offsets have been calculated:

Keep the sensor in a vertical position, as explained in Section 10.1. Attach weight fixtures
(brackets, hooks, etc), as required. Do not yet attach any weights or apply any calibration loads.
Then press ENTER.

6. The display will appear as follows:

Optionally exercise the sensor several times (at full scale, if possible), then press ENTER.

7. The display will appear as follows:


Apply torque equal to the full scale of the instrument, then press ENTER.

8. After displaying “PLEASE WAIT…” the display will appear as follows:

CALIBRATION

CLOCKWISE

GAIN ADJUST

APPLY FULL SCALE LOAD

50.000 LBFIN +/-20%

THEN PRESS ENTER

CALIBRATION

CLOCKWISE

ENSURE NO LOAD

THEN PRESS ZERO

CALIBRATION

CLOCKWISE

Attach necessary
weight fixtures.

THEN PRESS ENTER

CALIBRATION

CLOCKWISE

Optionally exercise
load cell a few times.

THEN PRESS ENTER

CALIBRATION

OFFSET


Sen.Offset Adj.Failed
Ana.Offset Adj.Failed

CALIBRATION

OFFSET

Sen.Offset Adj.Passed
Ana.Offset Adj.Passed

Advertising
This manual is related to the following products: