Cooper Instruments & Systems TT03C Digital Cap Torque Gage User Manual

Page 16

Advertising
background image

Series TT03 Digital Torque Gauges

User’s Guide


15

Remove the torque applied in Step 7, leave the fixtures in place, then press ZERO.

9. The display will appear as follows:

Use the UP and DOWN keys to adjust the torque value as required. The torque values default to
even increments, as indicated by the previously entered number of data points described in Step
1. Then press ENTER.

Repeat the above step for the number of data points selected.

10. After all the clockwise calibration points have been completed, the display will appear as follows:

Press ENTER.

11. At the completion of the counter-clockwise calibration, the display will appear as follows:

To save the calibration information, select “SAVE & EXIT”. To exit without saving the data select
“EXIT W/O SAVING”.

12. Any errors are reported by the following screens:

Displayed at the start of calibration if a disallowed unit is selected.

CALIBRATION

Units must be kgFmm.

PLEASE TRY AGAIN

PRESS ENTER

CALIBRATION

COMPLETE


SAVE & EXIT

EXIT W/O SAVING

CALIBRATION

CLOCKWISE COMPLETE

REVERSE DIRECTION

FOR CCW

Attach necessary
weight fixtures.

THEN PRESS ENTER

CALIBRATION

CLOCKWISE

APPLY LOAD

1 OF 5

ENTER LOAD:

10.000 LBFIN

THEN PRESS ENTER