Calibration Error Messages

If unexpected pressures are encountered during the calibration procedure, the instrument will alert the user by
flashing the word "HELP” and a message number on the display. This indicates that the calibration process cannot
continue, and that you must turn the instrument off and re-calibrate again when the error has been corrected. A list
of error message numbers and their causes is given in "TROUBLESHOOTING" in Chapter 7.

CHAPTER 5

OUPUT SIGNALS

5.1

Output Signal Overview

Outputs of 4-20 mA, 0-5 VDC, and dual-Iimit relays are available as options. Each unit comes completely wired with
3 feet (standard) of cable so that the user can power and use the gauge immediately.

5.2 Specifications

5.2.1 Limit Relay Option

Power Required:

11-32 VDC @ 30 mA max.

Number Of Relays:

2 form "C" (i.e. normally open, common, normally closed). One
relay for "high" limit, another for "low" limit.

Connection:

Via terminal blocks inside case Interface: 3 feet of cable with
strain relief

Update Rate:

same as display

Relay Contact Rating:

1 Amp @ 24 VDC
0.5 Amp @ 48 VDC
Max. switched power: 24W
Max. switched voltage: 60V peak-peak

Limit Setup:

Via front panel push buttons

Limit Hysteresis:

1% of full scale

Limit Indicators:

Front panel LED
"L 1" LED is lit when pressure is below the low limit setpoint.
"L2" LED is lit when pressure is above the high limit setpoint.


5.2.2 0-5 VDC Output

Power Required:

11-32 VDC @ 30 mA max.

Output:

0-5 VDC +1- 0.25% of full scale output

Display & Output Zero Adjust:

Via front panel push buttons

Output Only Zero Adjust:

Internal Potentiometer

Output Only Span Adjust:

Internal Potentiometer

Operating Temperature:

30 to 160 degrees F

Connection:

Via terminal blocks inside case

Interface:

3 feet of cable with strain relief

Update Rate:

same as display

Max. Output Current:

0.2 mA

Short Circuit Protection:

+Output to -Output

Minimum Output:

At least -2.5% of full scale

Maximum Output:

At least 102.5% of full scale

Reverse Voltage Protected:

Yes

5.2.3 2-wire 4-20 mA Output

Supply Voltage:

2-wire loop, 11-32 VDC (load dependent, see figure 4-1)

Output:

4-20 mA +1- .25% of full scale output

Display & Output Zero Adjust:

Via front panel pushbuttons

Output Only Zero Adjust:

Internal Potentiometer

Output Only Span Adjust:

Internal Potentiometer

Operating Temperature:

30 to 160 degrees F

Connection:

Via terminal blocks inside case

DPG 100

11

Rev Feb 2001

Advertising