To change a feature with the Setup Menu:

Make sure the instrument is turned off. Press the [On/Off] button.

While unit is checking the display (lighting all LCD segments
simultaneously), press and hold down the [Zero] button.


The display now reads "L-LO”, which is the first item of the Setup Menu. Release the [Zero] button.

Pressing and releasing the [Zero] button will scroll down through the
available Setup Menu items.


The table below provides a list of the items available in the Setup Menu and a brief description of each.

Table 4: Setup Menu Items

Display Menu

Item Purpose

L-LO

low limit setpoint

Change low limit setpoint

L-HI

High limit setpoint

Change high limit setpoint

EO

Enable Options

Enable Auto-Off feature; Disable front
panel buttons

UNIT

Engineering units

Change engineering units used to
display pressure

A-LO

Analog Output Zero Scale

Force analog output to 0 Volts or 4
mA

A-HI

Analog Output Full Scale

Force analog output to 5 Volts or 20
mA

VER

Firmware version

Display internal firmware part number
and revision

When the menu item you wish to change is displayed, press [Clear].


The display now shows the present setting of that menu item. If you only wish to examine the present setting of the
menu item, you can use the [On/Off] button to turn the instrument off. Otherwise...

Use the [Hi/Lo] and [Clear] buttons to scroll up and down, respectively. Hold
a button down to change limit setpoints at a faster rate.


If you wish to abandon the changes you made to this setting, you can use the [On/Off] button to turn the instrument
off. Otherwise...

Once the setting you want is displayed, press the [Zero] button to store this
setting into memory.


After the instrument stores the setting into memory, the next menu item will be displayed.

DPG 100

16