8analog output zero scale ("a-lo") item, 9analog output full scale ("a-hi ") menu item, 10internal firmware version ("ver") menu item – Cooper Instruments & Systems DPG 100 Digital Pressure Test Gage User Manual

Page 22: Chapter 7troubleshooting, 1introduction, 2help message codes, Analog output zero scale (" a-lo ") item, Analog output full scale (" a-hi ") menu item, Internal firmware version (" ver ") menu item, Chapter 7 troubleshooting

Advertising
background image

Table 7: Available Units for "M" Pressure Reference

Setting

Engineering Unit I Comment

ADC

for factory use only

PSI

PSI

DUAL

inHg {<0 PSI) or PSI > 0 PSI)


6.8

Analog Output Zero Scale ("A-LO") Item

The Analog Output Zero Scale ("A-LO”) menu item forces the analog output to either 0 Volts or 4 mA, then displays
"DONE" on the display.

For more information on trimming the analog output, see "Trimming Of Analog Outputs" in Section 5.6.

6.9

Analog Output Full Scale ("A-HI ") Menu Item

The Analog Output Full Scale ("A-HI”) menu item forces the analog output to either 5 Volts or 20 mA, then displays
"DONE" on the display.

For more information on trimming the analog output, see "Trimming Of Analog Outputs" in Section 5.6.

6.10 Internal Firmware Version ("VER") Menu Item

The Internal Firmware Version ("VER") menu item displays the part number and version number of instrument's
operating program. The firm-ware part number and version number are of the form:

084-1087-011.00

Where the "084-1087-01" is the part number, and the "1.00" is the version number.

Since this information is too long to fit on the 4 digit-display, pressing either the [Hi/Lo] or [Clear] buttons will scroll
through this information 4 characters at a time.

CHAPTER 7

TROUBLESHOOTING

7.1 Introduction

This chapter provides information on correcting common problems that may be encountered operating and
calibrating the instrument.

7.2

Help Message Codes

If the instrument detects a problem during its power-on self-test, operation, or calibration, it will alert the user by
flashing the word "HELP" and an error message code number on the display. The instrument cannot continue
operation and you must turn the instrument off and correct the error.

"HELP 01”: Calibration error.
Analog to digital converter overrange.
One of the pressure points is above the calibration range of the instrument, or the sensor cable is not connected
properly.

"HELP 02”: Calibration error.
Analog to digital converter underrange.
One of the pressure points is below the calibration range of the instrument, or the sensor cable is not connected
properly.

"HELP 04” : Calibration error.
The applied pressures at any two calibration points did not differ enough.

DPG 100

19

Rev Feb 2001

Advertising