Model 5i Digital Force / Torque Indicator

User’s Guide


23

If failed:

5. The following screen appears after the offsets have been calculated:

Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply
any calibration loads. Then press ENTER.

6. The display will appear as follows:

Optionally exercise the sensor several times (at full scale, if possible), then press ENTER.

7. The display will appear as follows:

Apply a weight equal to the full scale of the instrument, then press ENTER.

8. After displaying “PLEASE WAIT…” the display will appear as follows:

CALIBRATION

COMPRESSION

GAIN ADJUST

APPLY FULL SCALE LOAD

10.000 LBF +/-20%

THEN PRESS ENTER

CALIBRATION

COMPRESSION

Attach necessary
weight fixtures.

THEN PRESS ENTER

CALIBRATION

COMPRESSION

Optionally exercise
load cell a few times.

THEN PRESS ENTER

CALIBRATION

OFFSET


Sen.Offset Adj.Failed
Ana.Offset Adj.Failed

CALIBRATION

OFFSET

Sen.Offset Adj.Passed
Ana.Offset Adj.Passed

CALIBRATION

OFFSET


Please wait…