3 mechanical setup – Cooper Instruments & Systems Model 5i Advanced Force/Torque Indicator User Manual

Page 6

Advertising
background image

Model 5i Digital Force / Torque Indicator

User’s Guide


5

The indicator can be configured to automatically power off following a period of inactivity. Refer to the
Other Settings section for details.

If battery replacement is necessary, the battery may be accessed by loosening the two captive screws in
the rear half of the housing and separating the two halves of the housing.

3 MECHANICAL SETUP


3.1 Connecting a sensor
The Plug & Test

TM

connector must be inserted into the receptacle of the 5i or 3i indicator with the side

marked “Plug & Test

TM

Technology” facing up (see Fig. 3.1). When fully inserted, the connector will lock

into place with a “click”.

Fig. 3.1
Appropriate orientation of Plug & Test

TM

connector. Sensor model

number, serial number, and load capacity may be found on the
labels affixed to the connector.


To release the connector, press both buttons on either side of the indicator housing to release the sensor
(see Fig. 3.2). Pull the connector completely out of the indicator by holding the curved aluminum section.
DO NOT pull on the cable or strain relief.

Fig. 3.2
Press both buttons on either side of the
indicator housing to release the Plug
& Test

TM

connector.


3.2 Sensor connector orientation
In order to accommodate a variety of testing requirements, the orientation of the Plug & Test

TM

connector

may be set up in either of the two positions shown below. To change the orientation, loosen the two
captive screws on the back side of the housing, separate the two housing halves, rotate one half 180
degrees, and reassemble. Contact between the two halves is made by the spring pins and contact pads
on the printed circuit boards.

Advertising