Current plot data can be exported to a selected tab‐delimited file for convenient importing in 
a spreadsheet or other application such as the VibraScout Post Processor.  This feature 
provides an alternative to the data log file by allowing the user to export data after 
acquisition. 
 
 

3. Device Info 

Model  #  and  Serial  #  are  stored  in  the  device  EEPROM  and  read  when  the  VibraScout 
application is launched.  Device memory is also available to store user information.  Enter up 
to 180 characters in the 

User Info

 field and click on 

Save User Info

 to store the information in 

non‐volatile memory.  When VibraScout is launched, the information will be read from the 
device and appear in the 

User Info

 field. 

 

4. Approximate Bandwidth 

Select  the  desired  bandwidth  based  on  the  application  requirements.    The  noise  level 
increases with bandwidth, so selecting a lower bandwidth will provide improved signal/noise 
at  the  expense  of  frequency  response.    The  7543A  sample  rate  is  equal  to  approximately 
twice  the  selected  bandwidth.    In  cases  where  data  overflows  occur  due  to  inadequate 
system resources (CPU speed, RAM), selecting a lower bandwidth will usually eliminate the 
overflows.  See the section on the 

Data Integrity

 indicator for more information. 

 

 

 
5. Data File Settings 

The 

Data File path

 is selected with the Browse button.  Data are saved as tab‐delimited text 

for  easy  importing  into  a  spreadsheet  or  other  application  when  the 

Log  To  File

  box  is 

checked  in  the  Acquisition  screen.  Software  automatically  appends  date  and  time  code  to 
the end of the file name for easier identification of test sequence. Data file is closed and this 
field is cleared every time acquisition is stopped.  Selecting the 

Log To File

 box will prompt 

user for creating a new data file. 

 

6. Offsets 

The  7543A  is  calibrated  at  the  Dytran  factory  along  each  orthogonal  axis  using  precision 
fixtures and earth’s gravity. When the 7543A is placed at any random position, the Offsets 

Acquire

 button allows the values for the three acceleration axes pitch and roll to be captured 

and  set  to  zero.    Therefore,  relative  measurements  may  now  be  made  from  that  position. 