COVIDIEN Genius™ 2 Tympanic Thermometer Checker-Calibrator User Manual

Page 11

Advertising
background image

Genius

TM

2

Checker/Calibrator

Table of Contents

9

Error 7: Wrong Target

This error is an indication that the Genius 2 Thermometer has been inserted into wrong target or was inserted into the target

prematurely. Please take care to follow the instructions on the screen and not perform steps prior to the instructions being

displayed.

The Error screen will be displayed for 60 seconds before returning to the Main Menu. To exit the Error screen sooner, press the

encoder knob.

Section VIII - Checker/Calibrator Specifications

Temperature Set Points

Low Target . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32.2 °C +/- 0.3 °C (90.0 °F +/- 0.5 °F)

High Target . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40.6 °C +/- 0.3 °C (105.0 °F +/- 0.5 °F)

Target Temperature Accuracy

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Initial Limit +/- 0.03 °C (0.05 °F)

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Post year 1 +/- 0.06 °C (0.10 °F)

Genius 2 Thermometer Accuracy After Recalibration*

Readings between 36. 0 °C and 39.0°C

(96.8 °F and 102.2 °F) . . . . . . . . . . . . . . . . . . . . . . . . +/- 0.2 °C (+/- 0.4 °F)

Readings less than 36. 0 °C (96.8 °F) or greater than 39.0°C (102.2 °F) . . . . . . . . . . . . . . . . . . . . +/- 0.3 °C (+/- 0.5 °F)

Warm-up Time Before Use

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15 minutes (from non-extreme ambient)

Ambient Temperature Operating Range

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.1-26.7 °C (70-80 °F)

Relative Humidity Operating Range

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 +/- 20%, non-condensing

Altitude Range

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Up to 2000 meters (6563 feet )

Storage Temperature Range

. . . . . . . . . . . . . . . . . . . . . . . . . -25 °C - 55 °C (-13°F - 131°F ) 85% RH non-condensing

Approximate Size

Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.1 cm (12.25” )

Depth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15.9 cm (6.25” )

Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26.0 cm (10.25” )

Weight

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3 kg (7.2 lbs )

Power Requirements

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100-240V 0.8-0.4A 47-63Hz

Fuse Requirements

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Schurter FSF 5 x 20mm Fast Blow

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Low Breaking Capacity 250 V, 3.15A

Recommended Checker/Calibrator Calibration Interval

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . One Year

ESD Immunity

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Meets IEC 61000-4-2 Non-continuous Criteria C

Pollution Degree

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

Installation Category

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Indoor Use Only

*Post recalibration accuracy using the Genius 2 Checker/Calibrator may not necessarily be equivalent to factory calibration

Section IX - Cleaning Instructions

Caution: Do not immerse the checker/calibrator, cables, or power cord in water or other cleaning solution; clean using

a damp (not wet) cloth. Failure to follow cleaning procedures described herein could result in hazards to users.
As with any AC powered electrical device, care must be taken to prevent liquid from entering the Checker/Calibrator to avoid

electrical shock hazard, fire hazard, or damage to electrical components.

Advertising