COVIDIEN Genius™ 2 Tympanic Thermometer Checker-Calibrator User Manual

Page 3

Advertising
background image

Genius

TM

2

Checker/Calibrator

Table of Contents

1

Section I - Functional Description

This manual describes the operation of the Genius™ 2 Checker/Calibrator. This device has been developed to check the accuracy

of the Genius™ 2 Tympanic Thermometer and automatically recalibrate the thermometer, if necessary. In addition, the Genius 2

Checker/Calibrator has the capability to generate a test report for each thermometer tested and save it to a USB flash drive.

The Genius 2 Checker/Calibrator will work with all Genius™ 2 Thermometer software revisions.

The Genius 2 Checker/Calibrator contains two independently controlled infrared calibration targets that are similar to factory

calibration targets. These “blackbody” targets are designed to have efficient radiative heat transfer and produce infrared radiation

that approximates the human ear at temperatures of 32.2° C (90.0° F ) and 40.6° C (105.0° F ).

F/C˚

F/C˚

1

2

3

4

8

9

10

5

6

7

11

F 3.15A 250V

Checker / Calibrator

Figure 1 - Front Panel Layout

Legend

1 Power Connector

7 High Temperature Target

2 Power Switch

8 High Temperature Target LED

3 Fuse Holder

9 LCD Display

4 Low Temperature Target LED

10 Encoder Knob

5 Low Temperature Target

11 USB port

6 Genius 2 Thermometer Connector

Advertising