10 general cleaning instructions, Some of the common problems are as follows, Genius – COVIDIEN Genius™ 2 Tympanic Thermometer Checker-Calibrator User Manual

Page 12: Checker/calibrator

Advertising
background image

Genius

TM

2

Checker/Calibrator

Table of Contents

10

General Cleaning Instructions

Cleaning of the Genius 2 Checker/Calibrator may be performed as follows:

First, unplug the Genius 2 Checker/Calibrator from power source. Never clean the device while connected to the

mains supply.

• A mild detergent should be used for general cleaning of outside surfaces. Also, a cloth damped with isopropyl alcohol can be

used to clean the unit. Do not use abrasives or solvents. Avoid excess moisture around the target assemblies and the electrical

connectors.

• Do not attempt to clean the target below the rubber flaps.

• Do not clean the target nests or flaps with alcohol.

If there is any doubt about the compatibility of a cleaning agent with part of this equipment or its materials contact

Covidien customer service (See Section XI).

Section X - Troubleshooting

Some of the common problems are as follows:

Problem

Probable Cause

Corrections

Genius 2 Checker / Calibrator will

not power up

Genius 2 Checker / Calibrator not receiving AC power

Check AC power connections

Blown fuse

Replace fuse

Test report does not write to USB

flash drive

USB flash drive not formatted

Use PC or laptop to format flash drive

USB flash drive write protected

Use PC or laptop to turn off write protection

USB flash drive does not contain enough free memory space

Use PC or laptop to delete unnecessary items or use different USB flash drive

USB flash drive incompatible with Genius 2 Checker/Calibrator

Use different USB flash drive

Genius 2 Checker / Calibrator has not had enough time to detect the

newly inserted USB flash drive

Retry write process

Multiple Genius 2 thermometers fail

recalibration

Probe cover stuck in one of the blackbody targets

Remove probe cover from blackbody target using small needle-nose pliers or

hemostats. Do not scratch the interior surface of the target.

Dirty probe cover

Use a new probe cover for each reading. Do not re-use a probe cover.

Probe covers should never be touched by hand.

Dirty thermometer lens

Clean the thermometer lens per the instructions in the Genius 2 operating

manual.

Genius 2 thermometer not seated correctly

Ensure that the Genius 2 thermometer is completely seated in the target and

at the proper angle.

Genius 2 thermometers, probe covers, and/or checker/calibrator not

at room temperature.

Ensure that the Genius 2 thermometers, probe covers, and/or checker/

calibrator have had enough time to equilibrate to room temperature before

proceeding.

Error 6 repeatedly appears

while testing the same Genius 2

thermometer

Potential power consumption issue with Genius 2 thermometer

Repeat calibration check / recalibration with a different Genius 2 thermometer

and return problem thermometer for factory servicing.

Genius 2 thermometer seems

to read high or low following a

successful recalibration

Probe cover reused during calibration check / recalibration

Repeat calibration check / recalibration using a new probe cover prior to every

target insertion during the process.