2. Slide the rotor onto the shaft and push down gently to seat the drive pins in the holes.

3. Place the retaining nut on the shaft and tighten securely using the 5/16 inch hex key.

You will not be able to tighten the retaining nut if the drive pins are not properly aligned. If this
occurs lift the rotor off and repeat the process until it is seated.

Spinning the Sample

CAUTION - Never operate the Express 3 without the rotor locked securely in place. Failure to
properly install and secure the rotor may result in damage to the centrifuge not covered by warranty.

Loading

1.

Ensure the appropriate tube inserts are installed. The Express 3 centrifuge is provided with
eight tube inserts for 13 x 75 mm tubes (Product No. SV08) and eight tube inserts for 16 x 100
mm tubes (Product No. SV07). Other sizes are available (See accessories).

2.

The rotor must be properly balanced to ensure smooth operation. DO NOT spin a single tube.

3.

Use the same size and style tubes in opposite positions. Balance liquid in tubes to within 0.5 mL.


CAUTION

- Replace Tube Insert immediately if a tube breaks during centrifugation.

CAUTION

- Replace tubes used for rotor balancing after 10 spin cycles to reduce chances of

breakage due to the repeated stress of centrifugation.

4.

Close and latch the centrifuge cover.

5.

Select the spin time by pressing the appropriate cycle time button.

6.

Press Start button.

Cycle Settings

The operator should select spin time settings to achieve optimum results for specific applications.
The following are general guidelines:

Setting

Speed

RCF

2 min

Produces high quality plasma/serum from whole

blood for chemistry, coagulation and cardiac

testing.

7200 rpm

4440 xg

3 min

Produces high quality plasma/serum in gel tubes

and large plastic collection tubes from whole blood

for chemistry or coagulation testing.

7200 rpm

4440 xg

5 min

Produces separation equivalent to 3 min setting.

Centrifugal forces are decreased to accommodate.

larger glass tubes which require lower g-forces.

5600 rpm

2685 xg

Tube Guide by Settings

Tube
Size

Insert Tube

Dimension

2 minute

3 minute

5 minute

5 mL

SV08

13x75 mm

Plastic

Plastic/Gel

Glass/Plastic/Gel

7 mL

SV06

16x75 mm

Plastic

Plastic/Gel

Glass/Plastic/Gel

7 mL

SV09

13x100 mm

Plastic

Plastic/Gel

Glass/Plastic/Gel

10 mL

SV07

16x100 mm

Plastic

Plastic/Gel

Glass/Plastic/Gel