11

Section

5

Appendix

Specifications

Product Number

SSX3

Model Number

M502-22

Spin Parameters

2 min: 120 seconds @ 7200 rpm / 4440 xg

3 min: 180 seconds @ 7200 rpm / 4440 xg

5 min: 300 seconds @ 5600 rpm / 2685 xg

Acceleration Time

≤ 25 seconds

Deceleration Time

≤ 30 seconds

Electrical

100 – 240 VAC, 50/60Hz, 3.0 Amps

Dimensions

Height: 7”

Width: 11”

Depth: 10”

Weight: 8¾ lbs.

Environmental

For indoor use only

Ambient temperature 2ºC to 40ºC

Maximum 80% RH between 2ºC and 30ºC,
decreasing to 50% RH at 40ºC

Maximum altitude 2000 m

Main supply voltage fluctuations should not exceed ±10%

Transient over-voltages according to installation category II

Pollution degree 2

Specifications are subject to change without notice.

WARNING – Do not operate or store the Express 3 or its rotor below the 2

o

C

minimum operating temperature. At extreme low temperature (below -18

o

C) the rotor

material is subject to cracking under impact stress (such as dropping onto a hard
surface).