How to use this manual

This manual along with information contained on product labels and in package inserts should
provide you with all the information you need to operate and maintain the Express 3.

Notes appear in italics to highlight information.

When the information requires special attention, a caution symbol appears next to the
italicized text.

Please pay close attention to the instructions that accompany the notes and symbols as well as the
standard laboratory practices outlined by your facility and local regulatory agencies. The table below
lists all the CAUTIONS and WARNINGS for the Express 3.

WARNING – Do not expose the Express 3 or its rotor to strong or concentrated
acids, bases or oxidizing agents, aromatic or halogenated hydrocarbons, esters or
ketones.

WARNING – Do not operate or store the Express 3 or its rotor below the 2

o

C

minimum operating temperature. At extreme low temperature (below -18

o

C) the

rotor material is subject to cracking under impact stress (such as dropping onto a
hard surface).

WARNING - Do not use the supplied line cord outside of North America. Use a
power cord with a grounded male connector suitable for the power outlet being
used, which is rated at 3.0 Amps or more and has an IEC320/CEE22 female
connector to attach to the centrifuge.

CAUTION – Never operate the Express 3 without the rotor locked securely in place.
Failure to properly install and secure the rotor may result in damage to the
centrifuge not covered by warranty.

CAUTION

- Replace Tube Insert immediately if a tube breaks during centrifugation.

CAUTION

- Replace tubes used for rotor balancing after 10 spin cycles to reduce

chances of breakage due to the repeated stress of centrifugation.

CAUTION- The rotor (RTX8) has a finite lifespan that is dependant on usage.
Cracks or swelling that develop indicate the rotor should be replaced immediately.

CAUTION

-

The cover interlock by-pass is for emergency use only. Misuse may

expose the operator to unsafe conditions.


CAUTION – Do not spray cleaning solutions directly onto the centrifuge bowl or
housing; overspray could reach the motor bearings or internal circuitry.
Subsequent damage may not be covered by warranty.

CAUTION

Unplug the Express 3 from the wall outlet before cleaning or

performing maintenance.

CAUTION – Follow Universal Precautions with all biological specimens, regardless
of whether the specimen is known to contain an infectious agent. (See references)

Please use the system as intended. Improper use of the Express 3 Primary Tube Centrifuge and/or
its accessories may cause damage to the system, inaccurate results, or potentially nullify warranties.

Do not operate or store instrument beyond specified temperatures.