Maintenance – Iris Sample Processing Statspin® MP Multipurpose Centrifuge User Manual

Page 12

Advertising
background image

8

Section

4

Maintenance

Overview

StatSpin recommends that instrument operators perform periodic inspections and preventative
maintenance on all StatSpin instruments. Contact StatSpin’s customer service department or
distributor if, at any time, the instrument is not functioning properly

.

CAUTION - Unplug the StatSpin MP from the wall outlet before performing
maintenance.


WARNING

-

Do not expose the StatSpin MP and its rotor to strong or concentrated

acids, bases, esters, aromatic or halogenated hydrocarbons, ketones or strong
oxidizing agents.

Cleaning

The outside surfaces and switch overlay panel can be cleaned with a water-dampened cloth and
mild detergent. The inner surface or bowl, a powder-coated steel surface, can be cleaned with a
mild detergent and disinfected if necessary by wiping with a cloth dampened with 70% alcohol or
10% bleach.

CAUTION - DO NOT spray the bowl or outer surfaces with detergent or bleach.
Excess liquid will harm the electronics and subsequent problems may not be covered
under warranty.

Checking the Rotor Speed

The rated speeds can be checked with a photoelectric tachometer available from many sources. If
the StatSpin MP fails to achieve operating speed (

+5%

) contact your distributor or StatSpin

Customer Service department.

Replacing the O-Ring

The figure on the left illustrates the position of the rubber O-Ring which is attached to
the rotor-holder. Should it ever break a new one can be installed as shown, by
weaving it behind and in front of the 6 pins on the rotor-holder.

The points at which the O-Ring touches the rotor are indicated by the letter “A”. Extra
O-Rings have been included.