How to use this manual

This manual along with information contained on product labels and in package inserts should
provide you with all the information you need to operate and maintain the StatSpin MP.

Notes appear in italics to highlight information. When the information requires special attention, a
caution symbol appears next to the italicized text.

Please pay close attention to the instructions that accompany the notes and symbols as well as the
standard laboratory practices outlined by your facility and local regulatory agencies. The table below
lists all the CAUTIONS/WARNINGS for the StatSpin MP.

WARNING - North American Installation: Only use the power supply included with the
unit. Use of other power supplies or transformers will damage the StatSpin MP
electronics and void the warranty.

WARNING - Outside of North America: Do Not Use the Line Cord Supplied. Use power
cord for at least 1.0 Amp or more with an IEC320/CEE22 female connector and male
connector suitable for the power outlet to be used.

CAUTION - The cover interlock bypass is designed for emergency use only. If the
equipment is not used properly, safety may be impaired.

CAUTION-. Do not leave any rotor on the rotor-holder when the StatSpin is not in use for
an extended period of time. Doing so may compress the O-Ring and decrease its ability
to hold rotors
.

CAUTION – Failure to properly install the rotor may result in damage to the centrifuge
and will void the warranty.

CAUTIONIf rotor is left in place between runs, be certain to “bottom” the rotor on the
rotor-holder before spinning another sample. Failure to properly “seat” the rotor each
time may result in the rotor becoming loose during centrifugation.

CAUTION - The RT12 and RM02 tube rotors must be balanced before operation. If only
one sample is being processed, a second sample can serve as the balance tube or use a
similar tube filled with water as a balance.


CAUTION DO NOT use glass tubes of any kind in the RT12 rotor.

IMPORTANT: All StatSpin rotors have a finite lifespan that is dependant on usage.
Rotors should be inspected for cracks and should be replaced immediately when any
crack or visible wear occurs.

CAUTION – Unplug the StatSpin MP from the wall outlet before performing maintenance.

WARNING: Do not expose the StatSpin MP and its rotor to strong or concentrated acids,
bases, esters, aromatic or halogenated hydrocarbons, ketones or strong oxidizing
agents.

CAUTION: DO NOT spray the bowl or outer surfaces with detergent or bleach. Excess
liquid will harm the electronics and subsequent problems may not be covered under
warranty.

CAUTION - Universal Precautions should be followed on all specimens, regardless of
whether a specimen is known to contain an infectious agent. (See references)

Please use the system as intended. Improper use of the StatSpin MP Multipurpose Centrifuge
and/or its accessories may cause damage to the system, inaccurate results, or potentially nullify
warranties.