Iris Sample Processing Statspin® MP Multipurpose Centrifuge User Manual

Page 13

Advertising
background image

9

Service

Refer all service to qualified service personnel. Reference the StatSpin Warranty for further
instruction. Be sure to complete and return the warranty card as directed.

Decontamination before returning for service

Any instrument or accessory containing accumulated blood and/or other biological or chemical
deposits must be cleaned prior to shipment to the manufacturer/dealer for service. This
decontamination is required by Federal Law (Title 48 and 49 of the Federal Regulations) and in
accordance with the Environmental Protection Agency’s Regulations for Biohazard Waste
Management. StatSpin personnel cannot perform this decontamination.

Troubleshooting

No LED’s (lamps) are on.

Is the power cord completely

plugged into the wall?

NO→

Plug power cord into the wall.

YES

Is the wall outlet working properly?

→ NO→

Try another wall outlet.

Is the barrel connector from the

transformer completely plugged into

the centrifuge?

→ NO→

Plug the barrel connector into the centrifuge.

YES

Contact StatSpin or an Authorized

Repair Facility.

Centrifuge Will Not Spin/Shuts

Off Prematurely

Is the cover completely closed and

cannot be opened manually?

NO→

Completely close the cover.

YES

Contact StatSpin or an Authorized

Repair Facility.

Centrifuge will not open at end

of cycle

Is the bowl gasket completely

seated and sealing the bowl?

NO→

Press the bowl gasket all the way down into
the housing slot.

YES

Contact StatSpin or an Authorized

Repair Facility.

Advertising