4.1 Input configuration

Selecting the input signal
Voltage ±200 VDC (*)
Voltage ±20 VDC (*)
Standard signal ±10 V
Standard signal ±20 mA
Shunt input ±100 mVDC

(*) Voltage ±200 VDC or ±20 VDC does not require further parame-

terization but device will directly skip to line „programming lock”.

4.2 Display configuration

Scaling mode
SCAL mode (keypad mode)
Teach-in mode

First measured value
Manual entry in SCAL mode.

In Teach mode, the value of the provided analog signal will be

adopted.

Assigning the first measured value a displayed value
Requires manual value entry and assignment using the keypad.

Admissible range:

from –1999 to 9999 with standard signals

from –1999 to 1999 with shunt input.

Decimal point
Position of decimal point (related to dSP1).

Second measured value
Requires manual entry in SCAL mode. In Teach mode, the value of

the provided analog signal will be adopted.

Assigning the second measured value a displayed value
Requires manual value entry and assignment using the keypad.

Admissible range:

from –1999 to 9999 with standard signals

from –1999 to 1999 with shunt input.