9troubleshooting – Baumer BA DenexEdgeDetectorII User Manual

Page 13

Advertising
background image

User Manual DENEX Edge Detector II

13/20

Baumer Electric AG

Version 2013-02, V1.6

www.baumer.com

Frauenfeld, Switzerland

9

TROUBLESHOOTING

The green LED (POWER) is OFF

The sensor is not connected to +24V on pin 1 and 0V on pin 6 on the Tuchel

connector.

If the sensor previously has been working, there could be transients damaging the

sensor. Make sure that the voltage is 24VDC and is well regulated.

The red LED (LASER-OFF) is lit constantly, and no laser is present

Re-power the sensor. Check that the laser is lit. Is the laser turned off after a while? If

so, the laser settings have changed and the sensor must be sent back for repair.

The orange LED (OUTPUT-signal) is lit for every copy, but no pulse, or a different number of
pulses, is received by the connected equipment

The interfacing is bad. Check the connection of the sensor. Is the sensor supposed to

source or sink the signal? Should the pulse be positive or negative? See chapter 3.3.

Is the totalizer used, a battery powered device? It can be a problem because of the

voltage drop over the output transistor in the sensor. There is approx. 0.8V voltage drop
over emitter and collector at 2 mA which could mean that the totalizer does not see any
changes in state, i.e. from ”high” to ”low”. Check with a scope between pulse and ground.

The opto-coupler in the sensor can be faulty.

There is an over count (too few copies in the bundles)

Check the mechanical installation. The sensor must be parallel to the stream.

Back-edges could be counted. Check by running one paper at a time, and see if

there is an extra pulse sent at the end of the paper.

If the surface of the product is wavy, try tilting the sensor 5 - 10

°

according to figure

below.

If there is a stitch or something near the edge, try a longer blocking or output pulse,
see chapter 4, ”DIP-SWITCH SETTINGS”

Advertising