Error warnings, Overview on commands, Crc error – Baumer N 140 User Manual

Page 17: Format error

Advertising
background image

N 140

www.baumer.com

17

5. Error warnings

5.1. CRC error

Upon recognizing a CRC error in a transmitted command the SPA will respond as follows:

SOH

Adr

Error

EOT

CRC

Response

01h

20h

65h

04h

46h Error = „e“ (65h) = CRC error


5.2. Format error

Upon recognizing a format error (incorrect length of protocol or void command (Cmd) in a transmitted command
the SPA will respond as follows:

SOH

Adr

Error

EOT

CRC

Response

01h

20h

66h

04h

40h Error = „f“ (66h) = Format error



6. Overview on commands

The following chart shows in alphabetical order all available commands (Cmd) as well as options regarding read
and write (parameterization).

Command

code

Data amount

in bytes

read

write

Broad-

cast

Save in

EEPROM

Significance

a (61h)

5

X

X

-

X

Read / write general parameters

b (62h)

8

X

X

-

X

Parameter: tolerance compensation, window

c (63h)

8

X

X

-

X

Parameter: scaling factor spindle pitch

i (69h)

1

X

X

X

X

Parameter: mm / inch

t (74h)

6

-

X

-

-

Transmit tool number

u (75h)

6

-

X

-

-

Transmit optional number sequence

x (78h)

5

X

X

-

X

Read / write specific parameters

A (41h)

2

X

X

X

X

Assign or show device identifier

C (43h)

3 / 11

X

-

-

-

Check if actual value = target

K (4Bh)

-

-

X

X

X

Specific command: Profile reset (delete)

Q (51h)

1

-

X

X

X

Specific command: N 140 reset

R (52h)

6

X

-

-

-

Read current value

S (53h)

8

X

X

-

X

Read/write target value

U (55h)

6

X

X

-

-

Offset

V (56h)

2

X

X

X

X

Read / write profile number

X (58h)

2 / 4 / 8

X

-

-

-

Specific command: read device data

Z (5Ah)

6

X

X

X

X

Set preset

Advertising