Initialization, Function: heating – Scilogex MS7-H550-Pro User Manual

Page 10

Advertising
background image

8

Digital 7” Square(Hotplate)Magnetic Stirrer

6.Initialization

6.1 Initializing sequence

Turn on instrument by the power switch.

The“Heat”and“Stir”LEDs and the LCD display
flash about 3 times.

LCD displays the work plate, safe temperature limit in

SAFE xxx℃”.

LCD displays residual heat warning“rES On/OFF”.

LCD displays stirring bar breakaway monitoring“br
On/OFF”.

LCD displays the“Probe”characters in case of
connecting with external temperature sensors.

6.2 Default settings

The following table is supplied with default settings:

Items

Default settings

Setting temperature (℃ )

25

Setting rotary speed (rpm)

100

Table 4

7.Function: Heating

The device is controlled by digital temperature control

technology, which has two separate safe circuits. The

hotplate is kept at a constant temperature by a digital control

circuit. The hotplate temperature can also be monitored

from a separate, adjustable safe circuit. The two temperature

sensors (PT1000) internal for temperature control are built

into the hotplate. The single external PT1000 can monitor

the temperature of sample.

If external temperature sensors needed, it must be plugged in

before the device is switched ON.

Ensure the external temperature sensor connect firmly
before heating.

Set the temperature via turning the temperature control
knob slowly to the ideal value.

When the heating function is switched on, the LED

Heat”will light up and the LCD will display the real

temperature.

The rated temperature will be displayed on the right-hand
side of the LCD as well as Temp and℃ characters.

The heating function is switched on or off by pushing
heating knob.

Instrument is turned on for the setting zone temperature

setting when the last shutdown temperature, the common