Digital 7” Square(Hotplate)Magnetic Stirrer

Manufacture’s standard accessory. If the sensor is not

plugged in,“Probe”will flash. If plugged in,“Probe”

will be shown on the digital display to indicate the sensor is

operating. The setting value of external temperature sensors

and actual temperature are displayed. Safe circuit controls

hotplate temperature.

Comparing with the temperature control of the hotplate,

the external temperature sensor can control the medium’s

temperature more precisely.

Warning!
Do not insert or remove sensor while heating.

7.2 Residual heat warning (Hot)

In order to prevent the risk of burns from a hotplate, digital

hotplate model has a residual heat warning function.

When the heating function is switched off and the hotplate

temperature is still above 50℃,“Hot”will flash to warn

that there is a hazard of burns from the hotplate (Figure

6).When the hotplate temperature drops to below 50℃, the

unit will automatically switch off. If users want to turn off

the LCD immediately, just pull out the plug directly. When

the plug is pulled out, the residual heat warning function can

not be run.

case, set the heating temperature display the actual

temperature may have the following differences:.

-

Hotplate center and outer edge.

-

The sample container and the container.

In order to ensure the accuracy of the temperature inside

the container, please use the external temperature sensor

PT1000.

7.1 Working with external temperature

sensor

Figure 5

The external temperature sensor PT1000 is the