Trial run – Scilogex MS7-H550-Pro User Manual

Page 9

Advertising
background image

7

Digital 7” Square(Hotplate)Magnetic Stirrer

Remote

Display in case of remote control.

Err

Display in case of error happening.

Rated value/Real

value

Display value in case of heating or

stirring function switching ON.

Table 3

Note:

If both heating and stirring functions have been

started at the same time, heating function

always has higher priority. If in this case speed

is changed via the stirring knob, it displays

stirring speed and reverses to temperature in

the duration of 5 seconds.

5.Trial run

Make sure the required operating voltage and power
supply voltage match.

Ensure the socket must be properly grounded.

Ensure the power is off.

Plug in the power cable,ensure the power is on and begin
initializing.

Add the medium into the vessel with a stirring bar.

Place vessel on the work plate.

Set the rated stirring speed and begin.

Observe the stirring bar and LCD display.

Set the rated temperature and start heating.

Observe the real temperature on LCD display.

Stop the heating and stirring functions.

If these operations above are normal, the device is ready

to operate. If these operations are not normal, the device

may be damaged during transportation, please contact

manufacturer for technical support.

Advertising