0 calibrating the model 3301c, Calibration steps – ETS-Lindgren 3301C Active Monopole User Manual

Page 43

Advertising
background image

7.0 Calibrating the Model 3301C

Before connecting any components, follow the

safety information in the ETS-Lindgren

Product Information Bulletin included with your

shipment.

The ETS-Lindgren 3301CAL

Calibration Fixture

incorporates

a resistive T network and a
10-picofarad capacitor to perform
the Equivalent Capacitive
Substitution Method (ECSM) of
calibration for rod antennas. The T
allows for accurate reading of the
input to the 3301CAL, and the
capacitor feeds the amplifier
through the same impedance as
the rod.

Calibration Steps

Depending on the your calibration configuration, see Model 3301C Calibration

With Receiver and Signal Generator

on page 46 or Model 3301C Calibration

With Network Analyzer

on page 47.

1.

Disconnect the Model 3301C Active Monopole Antenna from power.

2.

Remove the monopole element.

Calibrating the Model 3301C

| 43