General information

The device EasyFlow is delivered in a very useful carrying bag with shoulder strap, so it can be taken away on the

field in best conditions. It is recommended to replace the device and the probe in the bag after its use; this will
guaranty its protection. Although it has been designed to withstand rough working conditions, it is a high
precision instrument which should be protected from shocks. Used with a minimal of care your EasyFlow will be
reliable and long lasting work tool.

Technical characteristics

Device

Gauging range

0,1 l/s to 99'900 l/s

Gauging accuracy

< 5 % (with optimum mix of tracer)

Gauging repeatability

± 1%

Tracer type

Cooking salt (NaCl)

Tracer quantity

10 g to 100 kg of salt

Tracer ideal mix

Between 5 and 20 g of salt per l/s of estimated flow. (Ex : ~300 l/s -> 2 kg)

Supply

3 x 1,5V alkaline batteries, AA, LR6 type

Autonomy

About 100 hours under normal conditions

Communication

Serial link RS-232

Dimensions / weight

Complete bag : 260 x 190 x 130 mm / 1240 g
Device with probe : 230 x 150 x 80 mm / 620 g

Waterproof

IP65

Salinity

Measuring range

0 to 3200 mg/l

Sensitivity

1 mg/l

Precision

< 1 %

Temperature

Measuring range

0 to +40 °C

Precision

± 0,2 °C

Switch on the device

When pressing the ON/SELECT button the EasyFlow is switched on and this display
(on the right) appears after 2 seconds. It contains the version of the device and
actual batteries tension.


If this tension is under 3.3V this message appears and the batteries should be

replaced. Press ENTER to continue.


If the tension is under 3V, this message is shown during 5 seconds, then the device
shutdown. It is here necessary to change the batteries before working with the
EasyFlow. (See previous chapter)