Calibration, Calibration 8 – KMC Controls SAE-1162 User Manual

Page 8

Advertising
background image

SAE-1111/1112/1162 Carbon Monoxide (CO) Detectors

8

Installation Guide

NOTE: The sensor must be continuously powered

for at least 1/2 hour prior to calibration.

NOTE: If the “0 Filter” option is enabled, disable it

in the menu temporarily for the calibration.

(See

Configuration on page 4

.)

NOTE: The ZERO and SPAN potentiometers are

located on the edge of the sensor board.

NOTE: Attach the tubing as described and

illustrated above.

NOTE: Properly recalibrating the SAE-1162 duct

sensors shipped before August 2014 is more

difficult. For those earlier units with different

sensor boards (without the gas tube port

receptacle), it is recommended to disable

the ReCal mode, set the Fault Mode time to

3 years, and then replace the SAE-1162 with

a new or factory-recalibrated unit when the

time has expired.

1. Calibrate the sensor first in clean air with no CO

gas present by adjusting the ZERO potentiometer

on the sensor board until a 4 mA (or 0 VDC) out-

put is obtained and the LCD displays 0 ppm.

2. Turn the regulator on/off knob fully off, attach it

to the 250 ppm gas bottle, and firmly tighten the

regulator by hand.

3. Moisten the sponge and squeeze out any excess

water. Place the sponge in the cap so that it will

not touch the sensor but does not plug the hole

in the side of the cap. (The sponge will ensure the

gas is in the right humidity range.)

4. Attach the cap to the port over the sensor. (If the

gas cap is too loose on the sensor port, wrap elec-

trical tape around the cap.)

5. Slowly turn the valve knob on the regulator to let

the gas begin flowing. The regulator will restrict

the flow rate to the specified 200 ml/min.

6. Wait for 5 minutes and then adjust the SPAN

potentiometer on the sensor board until the LCD

reads 250 ppm (or the CO concentration of the

gas cylinder).

7. Close the valve on the tank.
8. Calibration is complete. Reset the ReCal Mode if

it is enabled.

Calibration

The sensor may also be calibrated or verified with

CO gas as needed. This requires a field calibration

kit consisting of a bottle of gas (e.g., 250 ppm CO), a

tank pressure regulator with flow restrictor, and the

necessary tubing to supply the gas into the sensor.

Calibration can be done at 68–80° F (20–27° C). The

device cover must be opened to perform an actual

calibration.
For the SAE-1111/1112 space CO detectors, verifica-

tion (of the calibration) with gas can be done without

removing the device cover. Simply apply gas using

the calibration cap attached directly to the port on

the cover and monitor the output signal or LCD. For

the SAE-1162 duct sensor, open the cover to access

the port.

SAE-1111/1112

SAE-1162

(shipped after August 2014)

Advertising
This manual is related to the following products: