Rice Lake RL50210 TA, 50-500 lb. Capacities Mild Steel User Manual
Page 5
5
6.
Use only “hermetically sealed” load cells in washdown applications.
“Environmentally protected” load cells are not suitable for such applica-
tions and will be damaged. If tanks and surrounding equipment are
frequently steam cleaned, or if the load cell is subjected to direct washdown,
a protective shroud for the weighing assembly is recommended. Proper
drainage is nescessary so the weighing assembly is not standing in water.
7.
All support points should be equally stiff so that they deflect by the same
amount as the vessel is loaded.
2.2 Installing the RL50210 TA/65059 Mount
The RL50210 TA/65059 load cell mount installation instructions are presented in
the following paragraphs. Shown in the figure below are typical load cell mount
installations for the 50-250 and 500 lb. configurations.
50-250 lb
500 lb
NEOPRENE MOUNT
SOCKET HEAD
CAP SCREW
PLAIN WASHER
SOCKET HEAD
LOAD CELL SCREWS
LOCK WASHER
DEFLECTION
SPACER
CANTILEVER BEAM LOAD CELL
BASE PLATE
FLEX LOCK NUT
TYPE B
PLAIN WASHER
LOADING SPACER
TOP PLATE
NEOPRENE
MOUNT
SOCKET HEAD
CAP SCREW
LOCK WASHER
SOCKET HEAD
LOAD CELL SCREWS
LOCK
WASHER
DEFLECTION
SPACER
SHEAR BEAM LOAD CELL
BASE PLATE
MACHINE SCREW
OVERLOAD STOP
SOCKET HEAD
SET SCREW
NYLON PLUG
LOADING SPACER
FLAT WASHER
TOP PLATE
FLAT
WASHER
3