Bariatric Handrail Scale Technical Manual - Scale Configuration

9

5.0

Scale Configuration

Options and parameter setup are done through the scale configuration section and is used for setting values,

parameters, and options that are essential for the functioning of the scale. Access to the setup switch is located

under the tilt stand cover. Use a Phillips head screwdriver to remove the four screws holding the cover in place.

(shown in Figure 5-1— left photo).

Remove four screws to access setup switch

Setup switch location

B a t t e r y
compartment

Figure 5-1. Setup Switch Location

Use the following steps to enter into configuration mode.

1. Make sure the scale is turned off.
2. Turn the scale on by simultaneously pressing the

On/Off

key and the

ENTER

key.

3. Continue to hold both keys until

Id

appears. The unit cycles through its startup function and continues to

display the software version.

4. Access the recessed setup switch located in the back of the indicator. You can use a small paper clip,

small screwdriver or other similar object to press the setup switch.

5. Once the setup switch is pressed,

PROG

appears on the display.

6. The scale can be configured using a series of menus accessed through the front panel when the scale is in

setup mode.

DEF

CALIB

PROG

DONE

Figure 5-2. Top Level Menu

7. Press the

BMI

key to advance to the desired menu.

8. Press the

ENTER

key and advance in the manual to the related menu selection for further instructions.