Bariatric Handrail Scale Technical Manual - RS-232 Communication

19

7.2.2

Maintenance Protocol

Table 7-4 lists the maintenance protocol commands.

Table 7-4. Maintenance Protocol Commands

Command

Definition

R

Reboot

V

Firmware ID + development version

W

Current weight

A

Current AD

Z

Zero the scale

F

Show flash values (used for the first flash process)

L

USB On/Off (not available on USB communication

7.3

USB Connection

The Rice Lake Digital Bariatric Handrail Scale with seat has the capability of connecting to a PC using a USB

connection and a USB cable (not included). That connection location is shown in Figure 7-5.

USB Connection

RS-232 Connection

Load Cell Connection
Under This Cover

Figure 7-5. USB Connection Port

Connecting software and downloads should be addressed by your IT department and can vary depending on what

type of computer platform you’re using. Basic information on USB driver installation using Windows

®

is

described in the following steps and serves only as an example. The USB driver can be downloaded from the

Rice Lake Weighing Systems website at the following location; http://ricelake.com/software.aspx
Select Medical/Health Scales Downloads/Software and Get Downloads. Opening any product, will show a USB

Driver download. Click on

Download

to open and download the driver to your computer.

1. The graphic below shows the window that pops up when the USB cable is connected to the indicator and

the scale is turned on.