0 calibration, Calibration – Rice Lake BenchMark SLHE User Manual
Page 10
Advertising
6
Benchmark Series Single Point Bench Scales
3.0
Calibration
It is recommended that the scale be “exercised” by loading it to near capacity two or three times before calibration
to be certain that everything is seated. To calibrate your BenchMark:
1. With no load on scale, place the indicator in its calibration mode and perform a zero calibration.
2. Place test weights on platform equal to 70%–100% of scale’s capacity. If several weights are used, they
distribute them evenly around the platform.
3. Perform a span calibration.
4. Remove the test weights and check the zero reading. If necessary, repeat the calibration process.
See the indicator manual for the specific indicator calibration procedure.