18

Certificate Retrieval System User’s Guide

6.1.5

Test Result Screen

When the preload or final test readings have been
entered, this screen is displayed to select the pass/fail
result of the test. Each reading is displayed in the list
at the bottom, and the error is calculated.

Ensure the pass/fail reading is correct and click Next
to continue.

Figure 6-4. Test Result Screen

6.1.6

Weights Used

Once the preload and final tests are complete, the
weights used in the test must be selected. This screen
determines the traceability of the certification, and
care must be taken to select the right weights for the
test.

The left column lists the weight values and kits. Select
a value from this column to populate the center
column, which is displayed all weights in the system
of the selected value.

To assign weights to the certificate, highlight the
weights in the center column and click the right arrow
button. Weights are then added to the third column,
which should contain a list of all weights used on the
scale.

Figure 6-5. Weights Used Screen

6.1.7

Overall Results

This screen is used to set the overall results for the
certification. Select P (Pass), F (Fail), or A (Adjust)
from the dropdown.

If the scale is within the customer's application
requirement, check the Within Customer's Accuracy
checkbox.

F o r s c a l e s o r d e v i c e s c o n f i g u r e d f o r b a s i c
environmental conditions, a checkbox is displayed on
this screen for acceptable/unacceptable determination.
If advanced conditions are set, another screen will be
displayed to record environmental conditions.

Figure 6-6. Overall Results

6.1.8

Final Entry Screen

This screen is the last step in the calibration process.
Select the technician from the dropdown or type the
technician's name in the box. Select the location of the
test.

The Service ID can be recorded in the box or left
blank, and comments regarding the test can be entered
in the box provided.

To complete the calibration process, click the Next
button.

Figure 6-7. Final Entry Screen

6.1.9

Scale or Device Calibration Complete

Following a successful calibration, a pop-up window
is displayed which indicates the calibration record has
b e e n s a v e d . T h i s d i a l o g a l s o s h o w s t h e n e x t
calibration date for the scale. This is the date which
should be written on the scale calibration sticker.