2 balance service report – Rice Lake CRS-Certificate Retrieval System User Manual

Page 29

Advertising
background image

Certificate Retrieval System User’s Guide - Appendix

25

8.2

Balance Service Report

Form: Balance Sevice Report

Page 1 of 1

Customer:
Address:
City, State Zip:

Certificate ID:
ISO Number:
Date

123 Chemical Co.
P.O. Box 123 123 First Ave.
Atlanta, GEORGIA 30042

1231243124123092311
ISO/IEC 17025 A2LA #2039.01
11/23/2009

Indicator Mfg.

Indicator Model

Indicator Serial

Test Interval

Base Mfg.

Base Model

Base Serial

Cal Date

Due Date

Procedure

Scale ID

Scale Class

Scale Status

Scale Location

Scale Ranges

Sartorius

QC7DCE-S

1231231231231

30 Days

N/A

N/A

N/A

11/23/2009

12/23/2009

Sp.1

1231243124123

I

In Service

lab

0 - 1000 g x 0.00001 g
1001 - 2000 g x 0.00002 g
2001 - 3000 g x 0.003 g
3001 - 4000 g x 0.005 g

Customer Scale ID
xxx

Test
yes

17025 Compliant Test?
no

scale tolerance
yes

Weight value
yes

EQUIPMENT CONDITIONS

[Clean] Dirty Wet Really Drafty Hot Broken Not Found

As Found

Reading

Error

Section 1

Section 2

Section 3

Section 4

500.00000 g

500.00000 g

500.00000 g

500.00000 g

0.00000

0.00000

0.00000

0.00000

As Left

Reading

Error

Section 1

Section 2

Section 3

Section 4

500.00000 g

500.00000 g

500.00000 g

500.00000 g

0.00000

0.00000

0.00000

0.00000

Position

Position

1

2

4

3

1

2

3

4

SHIFT TEST

Shift Test Result: [Pass] Fail Adjust Not Applicable

Shift Weight: 500.00000 g

All tolerances calculated in conformance with Handbook 44 Table 6.

LOAD TEST

Preliminary Load Test

Test Wt.

Tolerance

Reading

Error

500.00000 g

1000.00000 g

2000.00000 g

3000.00000 g

4000.00000 g

500 - 500 g

1000 - 1000 g

2000 - 2000 g

2999.991 - 3000.009

3999.985 - 4000.015

500.00000 g

1000.00000 g

2000.00000 g

3000.00300 g

4000.00500 g

0.00000 g

0.00000 g

0.00000 g

0.00300 g

0.00500 g

Final Load Test

Test Wt.

Tolerance

Reading

Error

500.00000 g

1000.00000 g

2000.00000 g

3000.00000 g

4000.00000 g

500 - 500 g

1000 - 1000 g

2000 - 2000 g

2999.997 - 3000.003

3999.995 - 4000.005

500.00000 g

1000.00000 g

2000.00000 g

3000.00000 g

4000.00000 g

0.00000 g

0.00000 g

0.00000 g

0.00000 g

0.00000 g

In maintenance tolerance?

[Yes] No

In acceptance tolerance?

[Yes] No

TEST INFORMATION

Test Weight Classification:

F

Measurement Canada Traceability Certificate Number:

10333 Cal Date: 1/3/2008 Recal Date: 1/3/2008

Standards Used:

GRAM KIT

Test Location:

[Onsite] Offsite

Overall Result:

Pass Fail [Adjust] Unit Not Found Within Customers Tol

Was the scale within customers required accuracy?

[Yes] No Calibration Results

Environmental Conditions:

[Acceptable] Unacceptable

Comments / Notes:

gtg

Technician:

Brandon Cox

Uncertainty of Measurement

j

Clean Photoeye

y

Clean Infeed Roller

y

Devices were calibrated with the certified test weights. Adjustments made to restore and/or maintain the accuracy of the device conform to the
tolerances established by NIST or Manufacturers Specifications. Best measurement of uncertainty calculated using a coverage factor of K=2. This
provides confidence level of 95%.

Advertising