0 calibration – Rice Lake Digital Baby Scale (RL-DBS) - Technical and Operating Instructions User Manual

Page 9

Advertising
background image

Technical and Operating Instructions - Calibration

5

5.0

Calibration

To perform a calibration, connect the AC adaptor or install four (4) AA batteries and ensure that the scale is

sitting on a level surface.
The scale can be calibrated with either Lb or Kg weights. The unit is defaulted to calibrate with Lb weights so if

you are using LB weighs, proceed to Steps 1-7 (below). If you want to calibrate using Kg weights, perform Steps

A through D first and then proceed on to Steps 1-7.

A. Press and hold the

LB/KG

and the

ON/TARE

keys until the scale displays

CAL

. Release the keys and the

scale now displays the raw A/D counts.
B. Press the

RECALL

key and the display shows

CAL

with the LB indicator on. Press the

LB/KG

key to switch

between LB and KG and the display will now show

CAL

with the KG indicator on.

C. Press the

ON/TARE

key until the scale turns off.

D. Proceed to steps 1-7 shown in Table 5-1.
The following chart illustrates the calibration procedure.

Step

Display

1

To enter the Calibration Mode, press and hold the LB/KG
key and then press the ON/TARE key to turn on the scale.

- CAL -

then the raw A/D counts

2

Press the ON/TARE key again.

LB: Shows C20 w/ LB indicator symbol on
KG: Shows C10 w/ KG indicator symbol on

3

When the scale returns back to zero, load the scale with a
20 LB (10 KG if calibration is set to KG), test weight and
press the ON/TARE key.

LB: Shows C40 w/ LB indicator symbol on
KG: Shows C20 w/ KG indicator symbol on

4

When the scale returns back to zero, load the scale with a
40 LB (20 KG if calibration is set for KG), test weight and
press the ON/TARE key to finish the calibration.

- End -

5

Unit turns off automatically

6

Remove the 40 Lb (20 KG) weight from the scale.

7

Press the ON/TARE key to turn on the scale and place a
known weight onto the baby tray to confirm the correct
weight. If not, perform calibration again.

Table 5-1. Calibration Menu